{"cve":"CVE-2019-25718","nvd":{"published":"2026-06-01T23:16:13.270","modified":"2026-07-22T18:10:00.117","status":"Analyzed","description":"Dräger Infinity Explorer C700 contains a privilege escalation vulnerability that allows attackers to break out of kiosk mode and access the underlying operating system through a specific dialog interaction. Attackers can exploit this kiosk escape to take control of the operating system and cause the device to display incorrect or no information from the connected Delta Family patient monitor.","cvss_v4":{"vector":"CVSS:4.0/AV:L/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N","score":8.6},"cvss_v3":{"version":"3.1","vector":"CVSS:3.1/AV:L/AC:L/PR:L/UI:N/S:U/C:H/I:H/A:H","score":7.8},"cwes":["CWE-451"],"references":["https://static.draeger.com/security/download/2019-01-22-draeger-infinity-delta-vf10-1-security-advisory.pdf","https://www.vulncheck.com/advisories/dr-ger-infinity-explorer-c700-privilege-escalation-via-kiosk-mode-bypass"],"products":["draeger:infinity_explorer_c700_firmware"],"source":"nvd","synced_at":"2026-10-19T22:37:48.043517290+00:00"},"epss":{"cve":"CVE-2019-25718","score":0.00118,"percentile":0.01684,"date":"2026-10-31"},"kev":null,"osv":[],"attribution":["This product uses the NVD API but is not endorsed or certified by the NVD.","EPSS scores: FIRST Exploit Prediction Scoring System, https://www.first.org/epss","Known exploited vulnerabilities: CISA Known Exploited Vulnerabilities Catalog, https://www.cisa.gov/known-exploited-vulnerabilities-catalog","Package advisories: OSV (https://osv.dev), from the GitHub Advisory Database, PyPA, Go, RustSec and other ecosystem databases, under CC-BY 4.0 or CC0","CVE records: the CVE Program's CVE List (https://www.cve.org), under its terms of use (https://www.cve.org/Legal/TermsOfUse). CVE is a registered trademark of The MITRE Corporation; this product is not affiliated with the CVE Program."]}