{"cve":"CVE-2019-25716","nvd":{"published":"2026-06-01T22:16:17.170","modified":"2026-07-22T08:10:00.117","status":"Analyzed","description":"Dräger Infinity Delta, Delta XL, and Kappa patient monitors contain a denial-of-service vulnerability that allows remote attackers to cause the monitor to reboot by sending a malformed network packet. Attackers can repeatedly send malformed network packets to disrupt patient monitoring until the device falls back to default configuration and loses network connectivity.","cvss_v4":{"vector":"CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N","score":7.1},"cvss_v3":{"version":"3.1","vector":"CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H","score":7.5},"cwes":["CWE-15"],"references":["https://static.draeger.com/security/download/2019-01-22-draeger-infinity-delta-vf10-1-security-advisory.pdf","https://www.vulncheck.com/advisories/dr-ger-infinity-delta-kappa-patient-monitor-dos-via-malformed-network-packet"],"products":["draeger:delta_xl_firmware","draeger:infinity_delta_firmware","draeger:kappa_firmware"],"source":"nvd","synced_at":"2026-10-19T22:37:48.043517290+00:00"},"epss":{"cve":"CVE-2019-25716","score":0.00413,"percentile":0.35663,"date":"2026-10-31"},"kev":null,"osv":[],"attribution":["This product uses the NVD API but is not endorsed or certified by the NVD.","EPSS scores: FIRST Exploit Prediction Scoring System, https://www.first.org/epss","Known exploited vulnerabilities: CISA Known Exploited Vulnerabilities Catalog, https://www.cisa.gov/known-exploited-vulnerabilities-catalog","Package advisories: OSV (https://osv.dev), from the GitHub Advisory Database, PyPA, Go, RustSec and other ecosystem databases, under CC-BY 4.0 or CC0","CVE records: the CVE Program's CVE List (https://www.cve.org), under its terms of use (https://www.cve.org/Legal/TermsOfUse). CVE is a registered trademark of The MITRE Corporation; this product is not affiliated with the CVE Program."]}