{"cve":"CVE-2018-19014","nvd":{"published":"2019-01-28T22:29:00.397","modified":"2026-06-17T01:48:38.530","status":"Modified","description":"Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. Log files are accessible over an unauthenticated network connection. By accessing the log files, an attacker is able to gain insights about internals of the patient monitor, the location of the monitor, and wired network configuration.","cvss_v4":null,"cvss_v3":{"version":"3.0","vector":"CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N","score":6.5},"cwes":["CWE-532"],"references":["http://www.securityfocus.com/bid/106683","https://ics-cert.us-cert.gov/advisories/ICSMA-19-022-01","http://www.securityfocus.com/bid/106683","https://ics-cert.us-cert.gov/advisories/ICSMA-19-022-01"],"products":["draeger:delta_xl_firmware","draeger:infinity_delta_firmware","draeger:infinity_explorer_c700_firmware","draeger:kappa_firmware"],"source":"nvd","synced_at":"2026-10-19T20:54:46.948512734+00:00"},"epss":{"cve":"CVE-2018-19014","score":0.0078,"percentile":0.55924,"date":"2026-10-31"},"kev":null,"osv":[],"attribution":["This product uses the NVD API but is not endorsed or certified by the NVD.","EPSS scores: FIRST Exploit Prediction Scoring System, https://www.first.org/epss","Known exploited vulnerabilities: CISA Known Exploited Vulnerabilities Catalog, https://www.cisa.gov/known-exploited-vulnerabilities-catalog","Package advisories: OSV (https://osv.dev), from the GitHub Advisory Database, PyPA, Go, RustSec and other ecosystem databases, under CC-BY 4.0 or CC0","CVE records: the CVE Program's CVE List (https://www.cve.org), under its terms of use (https://www.cve.org/Legal/TermsOfUse). CVE is a registered trademark of The MITRE Corporation; this product is not affiliated with the CVE Program."]}