{"cve":"CVE-2018-19010","nvd":{"published":"2019-01-28T22:29:00.350","modified":"2026-06-17T01:48:38.050","status":"Modified","description":"Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. A malformed network packet may cause the monitor to reboot. By repeatedly sending the malformed network packet, an attacker may be able to disrupt patient monitoring by causing the monitor to repeatedly reboot until it falls back to default configuration and loses network connectivity.","cvss_v4":null,"cvss_v3":{"version":"3.0","vector":"CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H","score":6.5},"cwes":["CWE-20"],"references":["http://www.securityfocus.com/bid/106683","https://ics-cert.us-cert.gov/advisories/ICSMA-19-022-01","http://www.securityfocus.com/bid/106683","https://ics-cert.us-cert.gov/advisories/ICSMA-19-022-01"],"products":["draeger:delta_xl_firmware","draeger:infinity_delta_firmware","draeger:infinity_explorer_c700_firmware","draeger:kappa_firmware"],"source":"nvd","synced_at":"2026-10-19T20:54:46.948512734+00:00"},"epss":{"cve":"CVE-2018-19010","score":0.00762,"percentile":0.55366,"date":"2026-10-31"},"kev":null,"osv":[],"attribution":["This product uses the NVD API but is not endorsed or certified by the NVD.","EPSS scores: FIRST Exploit Prediction Scoring System, https://www.first.org/epss","Known exploited vulnerabilities: CISA Known Exploited Vulnerabilities Catalog, https://www.cisa.gov/known-exploited-vulnerabilities-catalog","Package advisories: OSV (https://osv.dev), from the GitHub Advisory Database, PyPA, Go, RustSec and other ecosystem databases, under CC-BY 4.0 or CC0","CVE records: the CVE Program's CVE List (https://www.cve.org), under its terms of use (https://www.cve.org/Legal/TermsOfUse). CVE is a registered trademark of The MITRE Corporation; this product is not affiliated with the CVE Program."]}