{"cve":"CVE-2018-10631","nvd":{"published":"2018-07-13T19:29:00.213","modified":"2026-06-17T01:34:19.063","status":"Modified","description":"The 8840 Clinician Programmer executes the application program from the 8870 Application Card. An attacker with physical access to an 8870 Application Card and sufficient technical capability can modify the contents of this card, including the binary executables. If modified to bypass protection mechanisms, this malicious code will be run when the card is inserted into an 8840 Clinician Programmer.","cvss_v4":null,"cvss_v3":{"version":"3.1","vector":"CVSS:3.1/AV:P/AC:H/PR:N/UI:R/S:U/C:H/I:H/A:H","score":6.3},"cwes":["CWE-693"],"references":["http://www.securityfocus.com/bid/104213","https://global.medtronic.com/xg-en/product-security/security-bulletins/nvision.html","https://ics-cert.us-cert.gov/advisories/ICSMA-18-137-01","https://www.medtronic.com/security","https://ics-cert.us-cert.gov/advisories/ICSMA-18-137-01"],"products":["medtronic:n\\'vision_8840_firmware","medtronic:n\\'vision_8870_firmware"],"source":"nvd","synced_at":"2026-10-19T20:52:03.425131697+00:00"},"epss":{"cve":"CVE-2018-10631","score":0.00411,"percentile":0.35322,"date":"2026-10-31"},"kev":null,"osv":[],"attribution":["This product uses the NVD API but is not endorsed or certified by the NVD.","EPSS scores: FIRST Exploit Prediction Scoring System, https://www.first.org/epss","Known exploited vulnerabilities: CISA Known Exploited Vulnerabilities Catalog, https://www.cisa.gov/known-exploited-vulnerabilities-catalog","Package advisories: OSV (https://osv.dev), from the GitHub Advisory Database, PyPA, Go, RustSec and other ecosystem databases, under CC-BY 4.0 or CC0","CVE records: the CVE Program's CVE List (https://www.cve.org), under its terms of use (https://www.cve.org/Legal/TermsOfUse). CVE is a registered trademark of The MITRE Corporation; this product is not affiliated with the CVE Program."]}