{"cve":"CVE-2017-12722","nvd":{"published":"2018-02-15T10:29:00.383","modified":"2026-06-17T01:03:50.690","status":"Modified","description":"An Out-of-bounds Read issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. A third-party component used in the pump reads memory out of bounds, causing the communications module to crash. Smiths Medical assesses that the crash of the communications module would not impact the operation of the therapeutic module.","cvss_v4":null,"cvss_v3":{"version":"3.0","vector":"CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:L","score":5.3},"cwes":["CWE-125"],"references":["http://www.securityfocus.com/bid/100665","http://www.securityfocus.com/bid/101252","https://ics-cert.us-cert.gov/advisories/ICSMA-17-250-02A","http://www.securityfocus.com/bid/100665","http://www.securityfocus.com/bid/101252"],"products":["smiths-medical:medfusion_4000_wireless_syringe_infusion_pump"],"source":"nvd","synced_at":"2026-10-19T20:48:34.567387330+00:00"},"epss":{"cve":"CVE-2017-12722","score":0.02415,"percentile":0.84139,"date":"2026-10-31"},"kev":null,"osv":[],"attribution":["This product uses the NVD API but is not endorsed or certified by the NVD.","EPSS scores: FIRST Exploit Prediction Scoring System, https://www.first.org/epss","Known exploited vulnerabilities: CISA Known Exploited Vulnerabilities Catalog, https://www.cisa.gov/known-exploited-vulnerabilities-catalog","Package advisories: OSV (https://osv.dev), from the GitHub Advisory Database, PyPA, Go, RustSec and other ecosystem databases, under CC-BY 4.0 or CC0","CVE records: the CVE Program's CVE List (https://www.cve.org), under its terms of use (https://www.cve.org/Legal/TermsOfUse). CVE is a registered trademark of The MITRE Corporation; this product is not affiliated with the CVE Program."]}