{"cve":"CVE-2015-3954","nvd":{"published":"2019-03-25T17:29:00.670","modified":"2026-06-17T00:26:32.110","status":"Modified","description":"Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges on Port 23/TELNET by default. An unauthorized user could issue commands to the pump. Hospira recommends that customers close Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.","cvss_v4":null,"cvss_v3":{"version":"3.0","vector":"CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H","score":9.8},"cwes":["CWE-285"],"references":["https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01","https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01"],"products":["pifzer:plum_a\\+3_infusion_system_firmware","pifzer:plum_a\\+_infusion_system_firmware","pifzer:symbiq_infusion_system_firmware"],"source":"nvd","synced_at":"2026-10-19T20:55:22.313337056+00:00"},"epss":{"cve":"CVE-2015-3954","score":0.02036,"percentile":0.81093,"date":"2026-10-31"},"kev":null,"osv":[],"attribution":["This product uses the NVD API but is not endorsed or certified by the NVD.","EPSS scores: FIRST Exploit Prediction Scoring System, https://www.first.org/epss","Known exploited vulnerabilities: CISA Known Exploited Vulnerabilities Catalog, https://www.cisa.gov/known-exploited-vulnerabilities-catalog","Package advisories: OSV (https://osv.dev), from the GitHub Advisory Database, PyPA, Go, RustSec and other ecosystem databases, under CC-BY 4.0 or CC0","CVE records: the CVE Program's CVE List (https://www.cve.org), under its terms of use (https://www.cve.org/Legal/TermsOfUse). CVE is a registered trademark of The MITRE Corporation; this product is not affiliated with the CVE Program."]}